Wall Suction Regulators
Chemetron Vacutron - Continuous intermitten

Continuous Intermitten
SPECIFICATIONS
FEATURE
CONT/INT
Selector Switches - Modes Available : Regular Yes
Full line Yes
Off Yes
Operating Range: Regular 150 mmHg +/- 10
Full line 320 mmHg +/- 20
(Maximum vacuum levels are limited by the source vacuum available)  
Operation Line (Source) Vacuum Range 12" to 30" Hg Unit Calibrated at 18" Hg Vacuum Source
Overall Dimensions 6.8" (174mm) High X 2.65" (167mm) Wide x 4.07" (104mm) Deep
Weight 1.5 lbs. (0.75 kg)
Vacuum Gauge 0 to 300 mmHg
Range

(Increments of 5 mmHg)

Size 0 to 40 kPa (Increments of 2 kPa)

Type

Diaphragm

Materials of Construction: Case

Knobs

Internal Parts

Metal Parts

High Impact Polycarbonate

Soft Thermo Plastic Rubber Outer Shell Molded on Thermoplastic Body

Acetal Resin

Brass Stainless Steel and Plated Steel

Operating Temperature 35 Degrees F to 120 Degrees F
Tenacore - 647 E. Young Street, Santa Ana, CA 92705 - Phone: 714-444-4643 - Fax: 714-549-7835